April 27, 2015 5-17
Using the Nicolet system to setup and acquire EEG
Data catch up
(disconnected
less
than 10 minutes)
The following applies to an amplifier wired to the Nicolet acquisition station with no
wireless capability.
NOTE: If the amplifier is in range of a wireless access point, disconnecting the
amplifiers Ethernet cable will not result in an interruption of data to the Nicolet
acquisition station.
Scenario
The amplifier is acquiring data for a study and is connected either:
directly to a Nicolet acquisition station
- or -
to a network running the Nicolet acquisition station.
a. The Ethernet cable connecting the amplifier to the Nicolet acquisition station
is disconnected causing the Nicolet software to display an alert that a
disconnect has occurred.
b. The disconnect lasts less than 10 minutes. The default is 10 minutes.
c. The amplifier is reconnected to the network or to the same Nicolet
acquisition station to which it was initially connected.
d. After a moment, the Nicolet acquisition station displays that the amplifier
has been reconnected.
Do not stop acquisition from the Nicolet acquisition station until
data catchup is complete as indicated when the wireless panel displays ‘In
Range’ and the fast scrolling catchup eeg returns to live data.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the natus Nicolet EEGwireless32 and is the answer not in the manual?

natus Nicolet EEGwireless32 Specifications

General IconGeneral
Brandnatus
ModelNicolet EEGwireless32
CategoryAmplifier
LanguageEnglish

Related product manuals