April 27, 2015 4-7
Nicolet EEGwireless32/64 Operation
Stand Alone Mode feature (starting a recording from the amplifier)
You can now start a study from the amplifier without having a Nicolet brand system
attached. This study doesn't have patient identification until it is connected to a
Nicolet brand system at which time you can register the amplifier and assign patient
demographic information in NicVue. If the amplifier has not been connected to a
Nicolet brand system, the patient information appears as all zeroes. You can acquire
multiple studies from a single patient when connected to a Nicolet brand system, all
data will appear with appropriate start and stop times and dates marked.
NOTE: To acquire data from another patient, you must first connect to a Nicolet
brand system and download all stored data before beginning a recording on a
second patient. Only one study can be performed on the amplifier before data is
cleared from memory. See Clearing the amplifier’s onboard memory later in this
chapter.
1. Press the Start Recording button for 3 - 5 seconds to start a study.
2. The Start Recording LED illuminates when the study starts (Figure 1).
Figure 1
A Start Recording button
B Start Recording LED