Nicolet Systems
1-10 April 27, 2015
Inspecting the system
Routinely check the system for exterior damage. Periodically check the system
ground integrity, the system leakage current and the leakage current of the Nicolet
EEGwireless32/64 amplifier and accessories.
Cleaning instructions
We recommend that you consult with your hospital infection control department and
follow all appropriate policies. Routine equipment cleaning recommendations are
provided below.
Cleaning the Nicolet
EEGwireless32/64
amplifier
The Nicolet EEGwireless32/64 amplifier and accessories are classified as a
noncritical item for infection control purposes.
Do not use propanone (acetone) on any part of the equipment. No part of the system
may be autoclaved or sterilized by any means. According to the CDC, the proper
cleaning method for non-critical medical devices includes the following:
A456-N
A428-N
Cavacide
3MQuat
70% isopropyl alcohol.