LAMBDA-n
Lambda Select
Syntax:
LAMBDA-n <wavelengths
where n
is
either A or B
Parameters:
The parameter <wavelength>
is
of type <number> which rounds to an integer
with a range that depends on the detector being used. The units
of
<wave
length> are nanometers and should correspond to the wavelength of the light
source being measured.
Function:
This command
is
used to specify the wavelength
of
light being measured.
This will insure that the proper responsivity calibration point will be used
when the
2832-C
calculates measurement values.
Related Commands:
ATTN-n, ATTN-n?, LAMBDA-n?, RESP-n?, USRCAL-n,
USRCALn?, USRRESP-n, USRRESP-n?
LAMBDA-n?
Lambda Query
Syntax:
LAMBDA-n?
where n
is
either A or B
Parameters:
None
Function:
This query returns the wavelength that corresponds to the responsivity calibra-
tion point currently being used by the
2832-C
in measurement calculation.
Returns:
<wavelength>
<wavelength>
is
of
type <number> and represents
an
integer. The units
of
<wavelength> are nanometers.
Related Commands:
ATTN-n, ATTN-n?, LAMBDA-n, RESP-n?, USRCAL-n,
USRCALn?, USRRESP-n, USRRESP-n?
LLO
Local Lockout Enable
Syntax:
LLO
<states
Parameters:
The parameter <state>
is
either
0
or
1.
If
<state>
is
0
then local lock-out
is
disabled.
If
<state>
is
1
then local lock-out
is
enabled.
Function:
This command enables or disables the local lockout.
Returns:
None
Related Commands:
LLO?