STSIZEn?
Statistics Buffer Size Query
Syntax:
STSIZE-n?
where n is either
A
or
B
Parameters:
None
Function:
This query returns the statistics buffer size.
Returns: <size>
<size>
is
of type <number> and represents an integer in the range
0
to
100.
Related Commands: STMAX-n?, STMEAN-n?, STMIN-n, STMXMN-n?,
STSDEV-n?
,
STSIZE-n
STOP
Stop Both Channel Acquisition
Syntax:
STOP
Parameters:
None
Function:
This command terminates any acquisition of data currently in progress on
both channels.
If
data
is
not being acquired on a channel then this command
has no effect.
Related Commands: RUN, RUN-n, STOP-n
STOP-n
Stop Single Channel Acquisition
Syntax:
STOP-n
where n
is
either
A
or
B
Parameters:
None
Function:
This command terminates any acquisition of data currently in progress on the
specified channel.
If
data is not being acquired on that channel the command
has no effect.
Related Commands: RUN, RUN-n, STOP
STOREF-n
Store Reference Reading
Syntax:
STOREF-n
where
n
is either
A
or
B
Parameters:
None
Function:
This command takes the latest reading and stores it as a reference reading to
be used when making relative linear and dB measurements. The units of the
reference reading defaults to Watts.
Related
Commands:
REFSEL-n, REEFLn?, USRREF-n, USRREF-n?, STOREKn?