36
may take a while for the compartment to cool (at least 15 minutes to reach 4°C from room
temperature), which is why this is performed prior to setting up the testing procedure.
Figure 2.19 Autosampler Temperature menu.
9. Click “Next” to proceed to the next step.
10. The next screen displays a reset baseline graph for your reference [Figure 2.20]. The signal
will continue to baseline in the background while you set up your test.
Figure 2.20 Baseline screen.
11. Click “Next” again to proceed to the Test Setup.