Training Guide Course No. 196514
Level 2 Maintenance, X-1000 Series Dispensing Systems 2-36 P/N 196515 (Revision A)
Exercise Sheet 2-4-4
Fluid Air Pressure Gauge Configuration
1. Make sure the air pressure to the Valve 1 Fluid Air Pressure Gauge is OFF by
turning the Valve 1 Fluid Air Pressure Regulator knob to the right
(counterclockwise) until it stops.
2. To configure the Valve 1 Gauge, press and hold down the Set button for at
least 2 seconds
3. Select the Measuring Unit by pressing the Up button or Down button until the
desired measuring unit is displayed and then press Set.
Available units are kPa or MPa (PA), kgf/cm2 (GF), bar (bAr), and psi (PSi).
4. Verify the Output 1 type is set on the normally open (1no) setting. If it is not
displayed, press the Up or Down button until 1no is displayed and then press
Set.
5. Verify the Output 2 type is set on the normally open (2no) setting. If it is not
displayed, press the Up button or Down button until 2no is displayed and then
press Set.
The output function of the digital gauge is not used on your
dispensing system.
6. Verify the Response Time is set at 192 ms. If it is not displayed, press the Up or
Down button until 192 is displayed.
7. Verify the Setting Mode is set on the manual setting (nAn). If the manual
setting is not displayed, press the Up or Down button until nAn is displayed and
then press Set.
The gauge will now be in the pressure-measuring mode.
8. Zero the gauge by holding down both arrow buttons simultaneously until zeros
are displayed on the screen.
9. Set the Valve 1 fluid pressure to the desired level for your application by turning
the knob on the Valve 1 Fluid Air Pressure Regulator to the left (clockwise) until
the correct pressure is displayed on the screen.
To ensure accurate pressure measurement, allow the gauge to warm
up for 20 to 30 minutes before adjusting pressure level.
10. If applicable to your system configuration, repeat Steps 1 to 9 for the Valve 2
Fluid Air Pressure Gauge.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Nordson Asymtek X-1010 and is the answer not in the manual?

Nordson Asymtek X-1010 Specifications

General IconGeneral
BrandNordson
ModelAsymtek X-1010
CategoryDispenser
LanguageEnglish

Related product manuals