7-2
StatStrip Glucose Hospital Meter
3. Temperature - Meter will only work within the tem-
perature range of 59°F to 104°F (15°C to 40°C).
Return the meter to an environment within the
speciedtemperaturerangeof59°Fto104°F(15°C
to 40°C).
Figure 7.3 Analysis Error - Temperature Error Screen Alert
4. Bad Sample - Insert a new strip and rerun the test.
If the error code persists, perform the test using an
alternate test strip vial or alternate method.
Figure 7.4 Analysis Error - Bad Sample Screen Alert