7-3
7. TrShoot.
Instructions for Use Manual 7 Troubleshooting
5. Replace Strip - Occurs after insertion of strip or oc-
curs during analysis. Insert another strip and retest.
If the error code persists, perform the test using an
alternate test strip vial or alternate method.
Figure 7.5 Analysis Error - Replace Strip Screen Alert
6. Flow Error - The specimen was incorrectly drawn
intotheteststripduetoeitherinsufcientorincor-
rect sample application. Repeat the test with a new
strip. If the error code persists, perform test using
an alternate method.
Figure 7.6 Analysis Error - Flow Error Screen Alert
13:42
Remove strip to Continue
OK
Continue
Flow Error [C - #]
Op: ****
Analysis Error
Name: A Unit-1
13:42
Replace Strip [A - #]
OK
Continue
Remove strip to Continue
Op: ****
Analysis Error
Name: A Unit-1
StatStrip Glu IFU 1.86 PN55848E EN 510K.indd 3 9/13/18 8:29 AM