11-3-4 Center Prism Mount AB032900:
a)
Assemble the following Jigs and B2-BIC,
Align the scrcw having
circular
mark wi1.h
adjusting liolo.
I
B2-BI units
as
shown in Fig. 1 1-1 3:
B2KC1403: Alignment Jig
KN0005: Centering objective
C-1
:
1 Ox eyepiece with cross hair
(See Fig. 1 1-1 3)
11
3
5
Assemble procedure for BH2-BI
a)
Assemble B2-BI Unit to B2-B IC Unit with
four Screws
CUK3x6SA.
NOTE:
B2-B
I
1-licsc
two Units should be aligned flush
Fig.
11-13
'~~0005
Center
cross
l~ait
Fig.
11-14
witti each other at their adjoining edges.
I)) Wl1011 Alijlllment Jig B2KC1403
is
attached
to
1311%B1, three adjusting holes in
1371<C:1403
slio~~ld
be aligned with three
S1:1uws (:IJI<:3x13SA in Prism Mount
AL30329(10
of
l3H2
I31.
c)
lr~se~l
Eyepiece 10X with Cross Hairs
C-l
into the right eyepiece sleeve.
d)
By using the adjusting holes of the align-
ment jig, adjust position of the Prism
Mount
AB032900 until the cross hair of
C-1 intersects with the center of the
center cross hair.
(See Fig. 11-13 and Fig. 11-14)
e) If centration
is
impossible by adjusting
position of the Prism Mount, check Prism
P2 for
its
assembly. (See Fig. 11-1 1)
f) If P2 has no defect, adjust mounting
positions of both B2-BI Unit and B2-BIC
Unit.
g)
Insert C-1 into left Eyepiece sleeve, and
check that the deviation of the left
eyepiece axis
is
within two divisions, if
any, from the right eyepiece axis.
1 1-4-1 Assemble Graduated Plate
AB031900 by
using four Screws
PUK2x4SA.
(See Fig. 10-3)
11-4-2 Assemble Finger
Rest
L
AB032100 with
Finger
Rest
R
AB032200 by using Screw
PUK2xlOSA and two Screws CUK2.6x5SA
for each Finger
Rest.
(See Fig. 10-2)
11-4-3
Temporarily Clamp Diopter Ring
AB032700
by using three Screws ACU2.6x4SA.
(See Fig. 10-1)