Chapter 6 Compatible Reprocessing Methods
109
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
Reprocessing methods Suction
Valve
(MAJ-1443)
Air/water
valve
(MAJ-1444)
Cleaning
brush
(MAJ-1534)
Channel
plug
(MAJ-621)
Air/water
channel
cleaning
adapter
(MAJ-629)
Manual
cleaning
Alkaline enzymatic
detergent
190 times 190 times 100 times 190 times 190 times
Neutral enzymatic
detergent
190 times 190 times 100 times 190 times 190 times
Manual
disinfection
Peracetic acid 190 times 190 times 100 times 190 times 190 times
Glutaraldehyde 190 times 190 times 100 times 190 times 190 times
Drying Alcohol
300 times 300 times
Not
compatible
300 times 300 times
Automatic
cleaning
and
disinfection
AER ETD Double
(Peracetic acid)
190 times 190 times
Not
compatible
Not
compatible
190 times
OER-AW
*1
(Peracetic acid)
300 times 300 times
Not
compatible
Not
compatible
Not
compatible
WD (alkaline detergent,
thermal disinfection)
190 times 190 times 10 times 190 times 190 times
Sterilization Steam sterilization
(autoclaving)
50 times 50 times 10 times 50 times 50 times
Ethylene oxide gas Not
compatible
Not
compatible
Not
compatible
50 times 50 times
Table 6.19
*1 OER-AW is not available in the member states of the EU.