110
Chapter 6 Compatible Reprocessing Methods
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
Reprocessing methods Injection tube
(MH-946)
Suction
cleaning
adapter
(MH-856)
Cleaning
adapter for
instrument
channel port
(MAJ-350)
Washing tube
(MH-974)
Manual
cleaning
Alkaline enzymatic detergent 300 times 300 times 75 times 300 times
Neutral enzymatic detergent 300 times 300 times 75 times 300 times
Manual
disinfection
Peracetic acid 300 times 300 times 75 times 300 times
Glutaraldehyde 300 times 300 times 75 times 300 times
Drying Alcohol 300 times 300 times 75 times 300 times
Automatic
cleaning
and
disinfection
WD (alkaline detergent,
thermal disinfection)
150 times 300 times 75 times 300 times
Sterilization Steam sterilization
(autoclaving)
100 times 100 times 25 times 100 times
Ethylene oxide gas 25 times 25 times 25 times 25 times
Table 6.20