Chapter 9 Reprocessing the Endoscope (and related reprocessing accessories)
141
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
1. Turn the elevator control lever all the way in the opposite direction of the
U” direction. perform the following brushing in the detergent solution.
2. While holding the distal end, brush the groove of the interior of the forceps
elevator with the cleaning brush (MAJ-1534) until all debris is removed (see
Figure 9.9).
Figure 9.9
3. Look at the left side of the instrument channel opening from the distal end,
brush the interior of this part with the cleaning brush (MAJ-1534) until all
debris is removed (see Figure 9.10).
Figure 9.10
Cleaning brush
(MAJ-1534)
Forceps elevator
Cleaning brush
(MAJ-1534)