EasyManua.ls Logo

Olympus EVIS EXERA II - Page 148

Olympus EVIS EXERA II
262 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
142
Chapter 9 Reprocessing the Endoscope (and related reprocessing accessories)
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
4. Look at the right side of the instrument channel opening from the distal end,
brush the interior of this part with the cleaning brush (MAJ-1534) until all
debris is removed (see Figure 9.11).
Figure 9.11
5. Move the elevator control lever in the “ U” direction to raise the forceps
elevator.
6. While holding the distal end, brush both sides of the forceps elevator and
the opposite side of the groove of the forceps elevator with the cleaning
brush (MAJ-1534) until all debris is removed (see Figure 9.12).
Figure 9.12
7. Brush the distal end of the endoscope, using the cleaning brush
(MAJ-1534).
8. Move the elevator control lever all the way in the opposite direction of the
U” direction to lower the forceps elevator.
Cleaning brush
(MAJ-1534)
Cleaning brush
(MAJ-1534)
Forceps elevator

Table of Contents

Other manuals for Olympus EVIS EXERA II

Related product manuals