Chapter 9 Reprocessing the Endoscope (and related reprocessing accessories)
143
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
Flushing the interior of the forceps elevator
1.
Immerse the distal end in detergent solution, operate the elevator control
lever to raise and lower the forceps elevator 3 times.
2. With the forceps elevator raised, flush the interior of the forceps elevator
with detergent solution using the 30 cm
3
(30 ml) syringe (see Figure 9.13).
Figure 9.13
Brushing the channels
• Be sure to brush the inside of the instrument channel and
suction channel. Otherwise, insufficient cleaning and/or
disinfection of the endoscope may pose an infection control
risk to the patient and/or operators performing the next
procedure with the endoscope.
• To avoid splattering detergent solution when the single use
combination cleaning brush (BW-412T) is pulled out, keep
the endoscope submerged.