Chapter 9 Reprocessing the Endoscope (and related reprocessing accessories)
149
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
15. Reprocess the suction tube, the suction cleaning adapter and suction valve
as described in Section 10.1, “Summary of reprocessing the accessories”
on page 165.
Figure 9.15
Figure 9.16
Alcohol is flammable. Handle with care.
Suction cleaning
adapter
Weighted end
Suction
cylinder
Cleaning adapter for the
instrument channel port
Suction cleaning
adapter
Weighted
end
Suction
port
Instrument channel
Balloon channel
Endoscope connector
Universal cord
Suction connector
Suction tube
Suction
channel
Suction pump
Suction bottle
Instrument
channel port
Suction
valve