150
Chapter 9 Reprocessing the Endoscope (and related reprocessing accessories)
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
Flushing detergent solution into the air/water and balloon
channels
1.
Attach the channel plug’s biopsy valve cap to the cleaning adapter for
instrument channel port (see Figure 9.17).
2. Lower the channel plug’s cylinder plug onto the air/water and suction
cylinders and slide the plug until it stops (see Figure 9.17).
Figure 9.17
3. Attach the injection tube’s connector plug to the air and water supply
connectors on the endoscope connector (see Figures 9.18 and 9.19).
4. Attach the injection tube’s air pipe port to the air pipe on the endoscope
connector (see Figures 9.18 and 9.19).
5. Attach the injection tube’s suction channel tube to the suction connector on
the endoscope connector (see Figures 9.18 and 9.19).
6. Immerse the suction port of the injection tube in the detergent solution.
Figure 9.18
Air/water cylinder
Suction cylinder
Instrument
channel port
Biopsy valve cap
Cylinder plug
Cleaning adapter for the
instrument channel port
Suction port
Syringe
Air/water
channel tube
Air pipe port
Detergent solution
Connector plug
Suction channel tube