Chapter 9 Reprocessing the Endoscope (and related reprocessing accessories)
151
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
7. Attach the 30 cm
3
(30 ml) syringe to the injection tube’s air/water channel
port (see Figure 9.19).
8. Inject 150 cm
3
(150 ml) of detergent solution into the air/water channel and
the balloon channel.
9. Detach the channel plug, the cleaning adapter for instrument channel port
and injection tube from the endoscope, and leave all items immersed.
Figure 9.19
Channel plug
Suction
cylinder
Air/water cylinderAir channel
Water channel
Air/water nozzle Balloon channel
Universal cord
Suction port
Suction channel port
Syringe
Air/water
channel port
Injection tube
Air pipe
Air pipe port
Air supply connector
Water supply
connector
Connector plug
Suction connector