152
Chapter 9 Reprocessing the Endoscope (and related reprocessing accessories)
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
Flushing detergent solution into the elevator wire channel
1.
Attach the washing tube to the elevator channel plug (see Figure 9.20).
Figure 9.20
2. Use the 5 cm
3
(5 ml) syringe to inject 15 cm
3
(15 ml) of detergent solution
into the elevator wire channel via the washing tube (see Figure 9.21).
Figure 9.21
3. Detach the washing tube from the endoscope and immerse it in detergent
solution.
Elevator channel plug
Washing tube
Washing tube
Elevator channel
plug
5cm
3
(5 ml) syringe
Elevator wire channel
Endoscope connector