172
Chapter 10 Reprocessing the Accessories
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
2. If air bubbles exit, repeat Step 1.
Flush the injection tube (MH-946)
1. Attach the syringe to the suction channel port of the injection tube.
Figure 10.9
2. While immersing the syringe and the injection tube completely in the
detergent solution, slowly pull the syringe plunger to fill the syringe with the
detergent solution. Then push in the plunger to flush the suction channel
tube of the injection tube with 30 ml of the detergent solution. Confirm that
no air bubbles exit from the tube during the flush.
Figure 10.10
3. If air bubbles exit, repeat Step 2.
Syringe
Air/water channel port
Suction channel port
Syringe
Suction channel tube
Suction channel port