Chapter 10 Reprocessing the Accessories
171
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
2. If air bubbles exit, flush the part where air bubbles exit with 30 ml of the
detergent solution until no air bubbles exit.
Flush the suction cleaning adapter (MH-856)
1. While immersing the syringe and the suction cleaning adapter completely in
the detergent solution, slowly fill the syringe with the detergent solution,
press it to the connecting end of the suction cleaning adapter, and flush the
adapter with 30 ml of the detergent solution. Confirm that no air bubbles exit
from the adapter during the flush.
Figure 10.7
2. If air bubbles exit, repeat Step 1.
Flush the washing tube (MH-974)
1. While immersing the syringe and the washing tube completely in the
detergent solution, slowly fill the syringe with the detergent solution, press it
to the connecting end of the washing tube, and flush the tube with 30 ml of
the detergent solution. Confirm that no air bubbles exit from the tube during
the flush.
Figure 10.8
Syringe
Weighted end
Connecting end
Syringe
Connecting end
Luer port