Chapter 10 Reprocessing the Accessories
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
5. While immersing the air/water channel cleaning adapter completely in the
detergent solution and rotating the adapter 360 degrees, using the syringe,
flush the interior of seals and the back of skirt of the adapter thoroughly with
30 ml of the detergent solution each (4 areas). Confirm that no air bubbles
exit from the adapter during the flush.
Figure 10.5
6. If air bubbles exit, flush the part where air bubbles exit with 30 ml of the
detergent solution until no air bubbles exit.
Flush the biopsy valve (MAJ-853)
1. While immersing the biopsy valve completely in the detergent solution,
using the syringe, flush the interiors of the cap and the main body of the
valve from both sides with 30 ml of the detergent solution each (4 areas).
Confirm that no air bubbles exit from the valve during the flush.
Figure 10.6
Syringe
Skirt
Seal
Rotate
Rotate