Chapter 10 Reprocessing the Accessories
169
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
Flush the accessories with detergent solution
Flush the Air/water channel cleaning adapter (MAJ-629)
1. While immersing the air/water channel cleaning adapter completely in the
detergent solution, depress and release the piston of the adapter several
times. Confirm that no air bubbles exit from the adapter during these
actions.
2. If air bubbles exit, repeat Step 1 until no air bubbles exit.
3. While immersing the air/water channel cleaning adapter completely in the
detergent solution, using the syringe, flush the holes of the adapter with
30 ml of the detergent solution each (2 holes). Confirm that no air bubbles
exit from the adapter during the flush.
Figure 10.4
4. If air bubbles exit, flush the part where air bubbles exit with 30 ml of the
detergent solution until no air bubbles exit.