Chapter 10 Reprocessing the Accessories
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
Brush the biopsy valve (MAJ-853)
Use the channel cleaning brush part of the single use combination cleaning
brush (BW-412T).
Figure 10.2
1. While immersing the biopsy valve completely in the detergent solution,
brush the hole and the slit of the valve (2 areas).
Figure 10.3
2. Inspect whether there is debris on the bristles.
3. Clean the bristles in the detergent solution using your gloved fingertips to
remove any debris.
4. Repeat Step 1 through 3 until no debris is observed upon inspection of the
brush.
5. When using the single use combination cleaning brush, dispose of the
single use combination cleaning brush as described in Section 12.5,
“Disposal”.
Channel cleaning
brush part