EasyManua.ls Logo

Olympus EVIS EXERA II - Page 180

Olympus EVIS EXERA II
262 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
174
Chapter 10 Reprocessing the Accessories
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
7. While immersing the syringe and the injection tube completely in the
detergent solution, slowly fill the syringe with the detergent solution, press it
to one side of the connector plug’s opening, and flush the plug with 30 ml of
the detergent solution. Confirm that no air bubbles exit from another side of
the connector plug’s opening during the flush.
Figure 10.13
8. If air bubbles exit, repeat Step 7.
Flush the channel plug (MAJ-621)
1. While immersing the channel plug completely in the detergent solution,
using the syringe, flush the interiors, the springs, and the holes of the plug
with 30 ml of the detergent solution each (5 areas). Confirm that no air
bubbles exit from the plug during the flush.
Figure 10.14
Syringe
Injection tube
Connector plug
Syringe
Hole
Spring
Interior Hole

Table of Contents

Other manuals for Olympus EVIS EXERA II

Related product manuals