Chapter 10 Reprocessing the Accessories
175
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
2. If air bubbles exit, flush the part where air bubbles exit with 30 ml of the
detergent solution until no air bubbles exit.
Flush the cleaning adapter for instrument channel port
(MAJ-350)
1. While immersing the cleaning adapter for instrument channel port
(MAJ-350) completely in the detergent solution, using the syringe, flush the
interiors, the holes of the adapter with 30 ml of the detergent solution each
(2 areas). Confirm that no air bubbles exit from the holes during the flush.
Figure 10.15
2. If air bubbles exit, flush the part where air bubbles exit with 30 ml of the
detergent solution until no air bubbles exit.
Syringe