Chapter 10 Reprocessing the Accessories
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
3. While immersing the cleaning adapter for instrument channel port
(MAJ-350) completely in the detergent solution and rotating the adapter
360 degrees, using the syringe, flush the groove of the adapter thoroughly
with 30 ml of the detergent solution. Confirm that no air bubbles exit from
the groove during the flush.
Figure 10.16
4. If air bubbles exit, flush the part where air bubbles exit with 30 ml of the
detergent solution until no air bubbles exit.
Clean the cleaning brush (MAJ-1534)
1. While immersing the cleaning brush completely in the detergent solution,
clean the bristles of the brushes, using your gloved fingertips to remove any
debris.
Figure 10.17
2. Confirm that all debris is removed from the bristles.
Groove
Rotate