Chapter 10 Reprocessing the Accessories
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
Dry the channel plug (MAJ-621)
1. Feed compressed filtered air of 0.2 MPa or more, less than 0.5 MPa to the
interiors, the springs, and the holes of the channel plug each (5 areas) until
no alcohol or water exits from them.
Figure 10.33
2. Thoroughly dry the external surfaces of the channel plug by wiping with
sterile lint-free cloths.
Dry the cleaning adapter for instrument channel port
(MAJ-350)
1. Feed compressed filtered air of 0.2 MPa or more, less than 0.5 MPa to the
interiors, the holes of the adapter each (2 areas) until no alcohol or water
exits from them.
Figure 10.34
2. Thoroughly dry the external surfaces of the adapter by wiping with sterile
lint-free cloths.
Hole
Spring
Interior Hole