Chapter 10 Reprocessing the Accessories
187
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
3. Feed compressed filtered air of 0.2 MPa or more, less than 0.5 MPa from
the suction channel port and the air/water channel port each (2 ports) until
no alcohol or water exits from the suction channel tube, the air/water
channel tube, and filter tubes.
Figure 10.31
4. Remove the cloths from the injection tube.
5. Feed compressed filtered air of 0.2 MPa or more, less than 0.5 MPa from
the one side of the connector plug’s opening until no alcohol or water exits
from another side of the connector plug’s opening.
Figure 10.32
6. Thoroughly dry the external surfaces of the injection tube by wiping with
sterile lint-free cloths.
Air
Air
Suction
channel port
Air/water channel port
Filter tube
Air/water channel tube
Suction channel tube
Air
Injection tube
Connector plug