186
Chapter 10 Reprocessing the Accessories
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
Dry the injection tube (MH-946)
1. Cover the outlet of the suction channel tube, the air pipe port, the air/water
channel port and the suction channel port of the injection tube in sterile
lint-free cloths to prevent splashing from the openings of the injection tubes.
Figure 10.29
2. Feed compressed filtered air of 0.2 MPa or more, less than 0.5 MPa from
the suction ports of the injection tube each (2 ports) until no alcohol or water
exits from the filter tubes.
Figure 10.30
Filter tube
Suction port
AirAir