Chapter 10 Reprocessing the Accessories
185
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
Dry the washing tube (MH-974)
1. Cover the weighted end of the washing tube in sterile lint-free cloths to
prevent splashing from the weighted end of the tube.
Figure 10.27
2. Feed compressed filtered air of 0.2 MPa or more, less than 0.5 Mpa from
the luer port of the washing tube until no alcohol or water exits from the
tube.
Figure 10.28
3. Remove the cloths from the washing tube.
4. Thoroughly dry the external surfaces of the tube by wiping with sterile
lint-free cloths.