Chapter 10 Reprocessing the Accessories
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
Dry the suction cleaning adapter (MH-856)
1. Cover the weighted end of the suction cleaning adapter in sterile lint-free
cloths to prevent splashing from the weighted end of the adapter.
Figure 10.25
2. Feed compressed filtered air of 0.2 MPa or more, less than 0.5 MPa from
the connecting end of the suction cleaning adapter until no alcohol or water
exits from the adapter.
Figure 10.26
3. Remove the cloths from the suction cleaning adapter.
4. Thoroughly dry the external surfaces of the adapter by wiping with sterile
lint-free cloths.