Chapter 10 Reprocessing the Accessories
183
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
Figure 10.23
4. Thoroughly dry the external surfaces of the adapter by wiping with sterile
lint-free cloths.
Dry the biopsy valve (MAJ-853)
1. Feed compressed filtered air of 0.2 MPa or more, less than 0.5 MPa to the
interiors of the cap and the main body of the biopsy valve from both sides
each (4 areas) until no alcohol or water exits from the valve.
Figure 10.24
2. Thoroughly dry the external surfaces of the valve by wiping with sterile
lint-free cloths.