182
Chapter 10 Reprocessing the Accessories
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
7. Attach the syringe to the suction channel port of the injection tube (see
Figure 10.9).
8. Pull the syringe plunger to fill the syringe with air. Then push in the plunger
to flush the suction channel tube of the injection tube with air. Confirm that
no alcohol exits from the tube during the flush.
9. If alcohol exits, repeat Step 8.
10. Move the syringe to the air/water channel port of the injection tube (see
Figure 10.11).
11. Pull the syringe plunger to fill the syringe with air. Then push in the plunger
to flush the air/water channel tube of the injection tube with air. Confirm that
no alcohol exits from the air pipe port during the flush.
12. If alcohol exits, repeat Step 11.
Dry the accessories
Dry the Air/water channel cleaning adapter (MAJ-629)
1. While rotating the air/water channel cleaning adapter 360 degrees, feed
compressed filtered air of 0.2 MPa or more, less than 0.5 MPa to the button
of the air/water channel cleaning adapter until no alcohol or water exits from
the adapter.
2. While rotating the adapter 360 degrees, feed compressed filtered air of
0.2 MPa or more, less than 0.5 MPa to the back of skirt of the adapter until
no alcohol or water exits from the adapter.
3. Feed compressed filtered air of 0.2 MPa or more, less than 0.5 MPa to the
holes of the adapter until no alcohol or water exits from the adapter.
Figure 10.22