Chapter 10 Reprocessing the Accessories
181
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
Flush the injection tube (MH-946)
1. Completely immerse the suction port of the injection tube in the alcohol.
2. Attach the syringe to the suction channel port of the injection tube (see
Figure 10.9).
3. Slowly pull the syringe plunger to fill the syringe with the alcohol. Then push
in the plunger to flush the suction channel tube of the injection tube with
30 ml of the alcohol.
Figure 10.20
4. Move the syringe to the air/water channel port of the injection tube (see
Figure 10.11).
5. Slowly pull the syringe plunger to fill the syringe with the alcohol. Then push
in the plunger to flush the air/water channel tube of the injection tube with
30 ml of the alcohol.
Figure 10.21
6. Remove the suction port of the injection tube from the alcohol.
Suction port
Suction channel port
Syringe
Suction channel tube
Suction port
Air/water channel port
Air pipe port
Syringe