Chapter 10 Reprocessing the Accessories
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
Flush the suction cleaning adapter (MH-856)
1. Fill the syringe with the alcohol, press it to the connecting end of the suction
cleaning adapter, and flush the adapter with 30 ml of the alcohol.
Figure 10.18
2. Hold the adapter and tilt it to expel all alcohol remaining in the adapter.
Flush the washing tube (MH-974)
1. Fill the syringe with the alcohol, press it to the connecting end of the
washing tube, and flush the tube with 30 ml of the alcohol.
Figure 10.19
2. Hold the adapter and tilt it to expel all alcohol remaining in the tube.