Chapter 10 Reprocessing the Accessories
179
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
Dry the external surface
1. Remove all other accessories from the water.
2. Wipe and dry the external surfaces of all accessories, using clean lint-free
cloths.
3. Inspect all accessories for residual debris.
4. If debris is found on any accessory, return to the beginning of Section 10.2,
“Manually cleaning the accessories” and repeat the cleaning procedure until
all debris is removed.
(Optional) Alcohol flush
Preparation
Fill a sterile, small container with the alcohol as described in Section 6.7,
“Alcohol”.
Immerse the valves
1. Completely immerse the biopsy valve (MAJ-853), and the Air/water channel
cleaning adapter (MAJ-629) in the alcohol.
2. While immersing the Air/water channel cleaning adapter completely in the
alcohol, depress and release the pistons of the valves and the adapter
several times.
3. Remove the biopsy valve, and the Air/water channel cleaning adapter from
the alcohol.