EasyManua.ls Logo

Olympus EVIS EXERA II - Page 196

Olympus EVIS EXERA II
262 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
190
Chapter 10 Reprocessing the Accessories
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
3. While immersing the Air/water channel cleaning adapter completely in the
disinfectant solution, using the syringe, flush the holes of the adapter with
30 ml of the disinfectant solution each (2 holes). Confirm that no air bubbles
exit from the adapter during the flush.
Figure 10.35
4. If air bubbles exit, flush the part where air bubbles exit with 30 ml of the
disinfectant solution until no air bubbles exit.
5. While immersing the Air/water channel cleaning adapter completely in the
disinfectant solution and rotating the adapter 360 degrees, using the
syringe, flush the interior of seals and the back of skirt of the adapter
thoroughly with 30 ml of the disinfectant solution each (2 areas). Confirm
that no air bubbles exit from the adapter during the flush.
Figure 10.36
6. If air bubbles exit, flush the part where air bubbles exit with 30 ml of the
disinfectant solution until no air bubbles exit.
Syringe
Hole
Syringe
Skirt
Seal
Rotate
Rotate

Table of Contents

Other manuals for Olympus EVIS EXERA II

Related product manuals