Chapter 10 Reprocessing the Accessories
191
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
Flush the biopsy valve (MAJ-853)
1. While immersing the biopsy valve completely in the disinfectant solution,
using the syringe, flush the interiors of the cap and the main body of the
valve from both sides with 30 ml of the disinfectant solution each (4 areas).
Confirm that no air bubbles exit from the valve during the flush.
Figure 10.37
2. If air bubbles exit, flush the part where air bubbles exit with 30 ml of the
disinfectant solution until no air bubbles exit.
Syringe
Cap
Main body