Chapter 10 Reprocessing the Accessories
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
Flush the suction cleaning adapter (MH-856)
1. While immersing the syringe and the suction cleaning adapter completely in
the disinfectant solution, slowly fill the syringe with the disinfectant solution,
press it to the connecting end of the suction cleaning adapter, and flush the
adapter with 30 ml of the disinfectant solution. Confirm that no air bubbles
exit from the adapter.
Figure 10.38
2. If air bubbles exit, repeat Step 1.
Flush the washing tube (MH-974)
1. While immersing the syringe and the washing tube completely in the
disinfectant solution, slowly fill the syringe with the disinfectant solution,
press it to the luer port of the washing tube, and flush the tube with 30 ml of
the disinfectant solution. Confirm that no air bubbles exit from the tube.
Figure 10.39
2. If air bubbles exit, repeat Step 1.
Syringe
Weighted end
Connecting end
Syringe
Connecting end
Luer port