Chapter 10 Reprocessing the Accessories
195
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
7. While immersing the syringe and the injection tube completely in the
disinfectant solution, slowly fill the syringe with the disinfectant solution,
press it to one side of the connector plug’s opening, and flush the plug with
30 ml of the disinfectant solution. Confirm that no air bubbles exit from
another side of the connector plug’s opening during the flush.
Figure 10.44
8. If air bubbles exit, repeat Step 7.
Flush the channel plug (MAJ-621)
1. While immersing the channel plug completely in the disinfectant solution,
using the syringe, flush the interiors, the springs, and the holes of the plug
with 30 ml of the disinfectant solution each (5 areas). Confirm that no air
bubbles exit from the plug during the flush.
Figure 10.45
Syringe
Injection tube
Connector plug
Syringe
Hole
Spring
Interior Hole