196
Chapter 10 Reprocessing the Accessories
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
2. If air bubbles exit, flush the part where air bubbles exit with 30 ml of the
disinfectant solution until no air bubbles exit.
Flush the cleaning adapter for instrument channel port
(MAJ-350)
1. While immersing the cleaning adapter for instrument channel port
(MAJ-350) completely in the disinfectant solution, using the syringe, flush
the interiors, the holes of the adapter with 30 ml of the disinfectant solution
each (2 areas). Confirm that no air bubbles exit from the holes during the
flush.
Figure 10.46
2. If air bubbles exit, flush the part where air bubbles exit with 30 ml of the
disinfectant solution until no air bubbles exit.
Syringe