Chapter 10 Reprocessing the Accessories
197
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
3. While immersing the cleaning adapter for instrument channel port
(MAJ-350) completely in the disinfectant solution and rotating the adapter
360 degrees, using the syringe, flush the groove of the adapter thoroughly
with 30 ml of the disinfectant solution. Confirm that no air bubbles exit from
the groove during the flush.
Figure 10.47
4. If air bubbles exit, flush the part where air bubbles exit with 30 ml of the
disinfectant solution until no air bubbles exit.
Disinfecting the cleaning brush (MAJ-1534)
1.
While immersing the cleaning brush completely in the disinfectant solution,
rub the bristles of the brushes with your gloved fingertips. Confirm that no air
bubbles exit from the bristles during the rubbing.
Figure 10.48
2. If air bubbles exit, repeat Step 1.