Chapter 10 Reprocessing the Accessories
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
Immerse the accessories in disinfectant solution
1.
While immersing the accessory completely in the disinfectant solution,
detach the syringe from the accessory.
2. Confirm that all accessories are completely immersed in the disinfectant
solution.
3. While immersing the all accessories completely in the disinfectant solution,
wipe all external surfaces of them to remove air bubbles, using clean
lint-free cloths.
4. Confirm that there are no air bubbles on the surfaces of all accessories.
5. If air bubbles adhere to the surfaces, wipe them away using your gloved
fingertips or clean lint-free cloths.
6. Cover the basin of the disinfectant solution with a tight-fitting lid to minimize
the diffusion of disinfectant vapors.
7. Leave all accessories completely immersed in the disinfectant solution
during the recommended contact time according to the instructions of the
disinfectant manufacturer. Use a clock or timer to accurately measure the
disinfection contact time.
Remove the accessories from disinfectant solution
Remove the suction cleaning adapter (MH-856)
1. Remove the suction cleaning adapter from the disinfectant solution.
2. Hold the adapter and tilt it to expel the disinfectant solution remaining in the
adapter.
Remove the washing tube (MH-974)
1. Remove the washing tube from the disinfectant solution.
2. Hold the tube and tilt it to expel the disinfectant solution remaining in the
tube.