Chapter 10 Reprocessing the Accessories
199
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
Remove the injection tube (MH-946)
1. Remove the injection tube from the disinfectant solution.
2. Attach the syringe to the suction channel port of the injection tube (see
Figure 10.40).
3. Pull the syringe plunger to fill the syringe with air. Then push in the plunger
to flush the suction channel tube of the injection tube with 30 ml of air to
expel the disinfectant solution.
4. Move the syringe to the air/water channel port of the injection tube (see
Figure 10.42).
5. Pull the syringe plunger to fill the syringe with air. Then push in the plunger
to flush the air/water channel tube of the injection tube with 30 ml of air to
expel the disinfectant solution.
Remove all other accessories
Remove all other accessories from the disinfectant solution.