Chapter 10 Reprocessing the Accessories
203
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
Remove the washing tube (MH-974)
1. Remove the washing tube from the rinse water.
2. Hold the tube and tilt it to expel the rinse water remaining in the tube.
3. Place it in a sterile basin.
Remove the injection tube (MH-946)
1. Remove the injection tube from the rinse water.
2. Attach the syringe to the suction channel port of the injection tube (see
Figure 10.40).
3. Pull the syringe plunger to fill the syringe with air. Then push in the plunger
to flush the suction channel tube of the injection tube with 30 ml of air to
expel the rinse water.
4. Move the syringe to the air/water channel port of the injection tube (see
Figure 10.42).
5. Pull the syringe plunger to fill the syringe with air. Then push in the plunger
to flush the air/water channel tube of the injection tube with 30 ml of air to
expel the rinse water.
6. Place it in the sterile basin.
Dry external surface
1. Remove all other accessories from the rinse water. Place them in the sterile
basin.
2. Thoroughly dry the external surfaces of all accessories by wiping with sterile
lint-free cloths.
(Optional) Alcohol flush
Refer to “(Optional) Alcohol flush” on page 179 as the procedure is same.
Dry the accessories
Refer to “Dry the accessories” on page 182 as the procedure is same.