46
Chapter 3 Preparation and Inspection
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
11. Attach the suction tube from the suction pump to the suction connector on
the endoscope connector (see Figure 3.19).
Figure 3.19
Connection of the endoscope and ultrasonic cable
The electrical contacts inside the ultrasound connector have
sharp tips and may be damaged by handling. Do not touch
the electrical contacts.
For more information on combining the endoscope with the
ultrasonic cable, refer to the “System chart” in the Appendix.
When connect the endoscope-side connector of the
ultrasonic cable to the ultrasonic cable connector, do not
entwine the universal cord with the ultrasonic cable.
Equipment damage can result.
1. Make sure that the inside of the ultrasonic cable connector is dry and free of
debris.
2. Align the mark on the ultrasonic cable with mark A on the ultrasound
connector and push it until it stops (see Figure 3.20).
3. Turn the endoscope-side connector of the ultrasonic cable clockwise until it
stops (see Figure 3.20).
4. Confirm that the mark on the ultrasonic cable is aligned with mark B on the
endoscope connector (see Figure 3.20).
Suction pump
Suction connector
Suction tube