Chapter 3 Preparation and Inspection
51
ULTRASOUND GASTROVIDEOSCOPE GF-UCT180
Inspection of the instrument channel and forceps elevator
Keep your eyes away from the distal end when inserting
EndoTherapy accessories. Extending the EndoTherapy
accessory from the distal end could cause eye injury.
1. Confirm that the forceps elevator is lowered, then insert the EndoTherapy
accessory through the biopsy valve. Confirm that the EndoTherapy
accessory extends smoothly from the distal end, and that a foreign object
does not come out.
2. Extend the EndoTherapy accessory approximately 3 cm from the distal end.
Move the elevator control lever in the “ U” direction and confirm that the
forceps elevator is raised smoothly.
3. Move the elevator control lever in the opposite direction of the “ U”
direction and confirm that the forceps elevator is lowered.
4. Confirm that the EndoTherapy accessory can be withdrawn smoothly from
the biopsy valve.
Inspection of the ultrasound image with the Olympus
universal endoscopic ultrasound center EU-ME1
1.
Turn on Olympus universal endoscopic ultrasound center EU-ME1.
A progress bar is displayed after connecting the ultrasound
endoscope (see Figure 3.21). During displaying the progress
bar, the information on the connected ultrasound endoscope
has been updated to the ultrasound center. Do not turn OFF
the ultrasound center while the progress bar is displayed.
Figure 3.21