4.1 General flow of installation of the equipment
OER-Elite INSTALLATION MANUAL
Ch.4
WARNING
Do not push the rear end of equipment against a wall after connecting the power
cord plug. The power cord may break and result in an electric shock or a fire.
Make sure that the hospital-grade, wall mains power outlet of your medical facility
meets the power specifications required by this unit. Insufficient capacity could
result in a fire or activation of the facility’s circuit breakers, which will turn off the
power to all other equipment connected to the same power source.
Use the power cord plug supplied with the equipment. Using a different power cord
plug may result in equipment malfunction or power cord burnout. Additionally, do
not use the supplied power cord plug on any other equipment as it is specified to be
used only with this equipment.
CAUTION
To prevent fluid leakage, do not attempt to extend the water supply hose and drain
hoses.
To prevent fluid leakage or damage, do not bend the water supply hose and drain
hose or apply external force to them.
To prevent discharge failure, cut off any excess drain hose and coil in such a
manner as to prevent it from bending or kinking.
To prevent discharge failure, always leave the drain hose tip open to the air inside
the floor drain (the hose tip should not be immersed in liquid and the inside of the
floor drain should be open to the air).
If the water supply capacity drops below the minimum required flow rate and
pressure levels, the equipment may stop with an error, the process time may be
prolonged. When the water supply capacity pressure and flow rate is low, improve
the water supply by installing a booster pump, etc.
If the water supply pressure (shut-off pressure) exceeds 0.5 MPa, contact Olympus
to request pressure reducing materials.
Do not supply water with a temperature over 28qC (82qF), as this can cause the
equipment or endoscope to deteriorate.
Although the reprocessor incorporates a water filter, it may still fail or the water filter
use life may be reduced if the water contains a large amount of debris and
sediment. If this is the case, contact Olympus to have a strainer installed in the
water supply piping.