4.1 General flow of installation of the equipment
51
OER-Elite INSTALLATION MANUAL
Ch.4
CAUTION
The disinfectant solution’s waste fluid should be treated in accordance with the
instructions supplied with the disinfectant solution. It is recommended the waste
fluid be treated and drained or disposed of according to local wastewater standards
defined by law, or to temporarily collect and store the waste fluid and have it treated
by a waste disposal firm.
Do not install the equipment in a place exposed to direct sunlight or where the
temperature exceeds 40qC (104qF). Otherwise, the equipment may malfunction or
the disinfectant solution may not last as long.
Do not place objects on or apply excessive force to the equipment. Doing so could
damage or deform the lid.
Before installing the equipment, ventilate the room. Otherwise, the ancillary
equipment may malfunction.
Do not install equipment in a way that would make it difficult to disconnect the
power cord plug from the wall mains power outlet. Otherwise, power shutdown by
disconnecting the power cord plug cannot be performed in case of malfunction or
failure of the equipment.
Before installing the equipment, be sure to check that the installation meets the
conditions provided in Section 4.3, “Installation conditions”. If not, water leaks,
malfunction, or failure of the equipment may result.